Model: HNAG920
Sampling Method: Pumping Suction and Diffusion
Detection Principle:Electrochemistry
Gas: ozone (O3)
Range: 0-50PPM,0-100PPM.
Resolution: 0.01PPM,0.1PPM.
Accuracy: ≤±2% F.S.
Model: HNAG920
Sampling Method: Pumping Suction and Diffusion
Detection Principle:Electrochemistry
Gas: ozone (O3)
Range: 0-50PPM,0-100PPM.
Resolution: 0.01PPM,0.1PPM.
Accuracy: ≤±2% F.S.
● Equipped with a 32-bit microprocessor and high-speed, high-precision processing circuitry for fast and accurate measurements.
● Features an external USB port for data export and deletion via USB cable connection to a computer.
● Supports up to 5 different sensors for flexible detection.
● Built-in buttons for easy parameter viewing and configuration.
● Allows unit switching (e.g., ppm ↔ mg/m³) through coefficient adjustment.
● Customizable historical data storage intervals.
● Supports Chinese/English language display switching.
● Fall alarm / one-touch emergency alarm function.
● 180° auto-rotating screen.
● Temperature & humidity compensation.
● PC+TPU (rubber overmolding process) for impact resistance.
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
Gas Detection Q&A
1. On Calibration and Accuracy
Q: Why do detectors need regular calibration?
A: Sensors experience "drift" over time, leading to inaccurate readings. Regular calibration (typically recommended every 6 months or 1 year) is necessary to ensure data accuracy and compliance with regulatory requirements.
Q: What is the difference between zero calibration and span calibration?
A: Zero calibration adjusts the reading to zero in clean air to eliminate background interference; span calibration uses ozone gas of a standard concentration to verify the instrument's linear response, ensuring accuracy at high concentrations.
2. On Readings and Anomalies
Q: The instrument reading consistently shows "0" or extremely low values. Is it broken?
A: Not necessarily. First, check if the sensor is within its validity period (typically 1-2 years). Readings become invalid if the sensor has expired. Second, check if the "auto-zero" function is enabled, as it may automatically deduct background values at low concentrations.
Q: What should I do if the reading suddenly spikes or shows severe fluctuations?
A: Ensure your own safety first and quickly evacuate to a ventilated area. A spike could indicate a genuine high-concentration leak, or it could be due to sensor moisture, poisoning (exposure to high concentrations of interfering gases), or circuit failure. It is recommended to immediately cross-verify with another instrument.
3. On Maintenance and Care
Q: How often does the sensor need to be replaced?
A: The lifespan of electrochemical sensors is typically 2-3 years. Actual lifespan depends on the ozone concentration and temperature/humidity in the operating environment, as high-concentration environments accelerate sensor aging.
Q: How should I clean and store the instrument?
A: Use a soft, damp cloth to wipe the exterior. Do not use organic solvents (e.g., alcohol, acetone) to clean the sensor area. Store the instrument in a dry, cool place, avoiding direct sunlight and extreme temperatures.
● Equipped with a 32-bit microprocessor and high-speed, high-precision processing circuitry for fast and accurate measurements.
● Features an external USB port for data export and deletion via USB cable connection to a computer.
● Supports up to 5 different sensors for flexible detection.
● Built-in buttons for easy parameter viewing and configuration.
● Allows unit switching (e.g., ppm ↔ mg/m³) through coefficient adjustment.
● Customizable historical data storage intervals.
● Supports Chinese/English language display switching.
● Fall alarm / one-touch emergency alarm function.
● 180° auto-rotating screen.
● Temperature & humidity compensation.
● PC+TPU (rubber overmolding process) for impact resistance.






Gas Detection Q&A
1. On Calibration and Accuracy
Q: Why do detectors need regular calibration?
A: Sensors experience "drift" over time, leading to inaccurate readings. Regular calibration (typically recommended every 6 months or 1 year) is necessary to ensure data accuracy and compliance with regulatory requirements.
Q: What is the difference between zero calibration and span calibration?
A: Zero calibration adjusts the reading to zero in clean air to eliminate background interference; span calibration uses ozone gas of a standard concentration to verify the instrument's linear response, ensuring accuracy at high concentrations.
2. On Readings and Anomalies
Q: The instrument reading consistently shows "0" or extremely low values. Is it broken?
A: Not necessarily. First, check if the sensor is within its validity period (typically 1-2 years). Readings become invalid if the sensor has expired. Second, check if the "auto-zero" function is enabled, as it may automatically deduct background values at low concentrations.
Q: What should I do if the reading suddenly spikes or shows severe fluctuations?
A: Ensure your own safety first and quickly evacuate to a ventilated area. A spike could indicate a genuine high-concentration leak, or it could be due to sensor moisture, poisoning (exposure to high concentrations of interfering gases), or circuit failure. It is recommended to immediately cross-verify with another instrument.
3. On Maintenance and Care
Q: How often does the sensor need to be replaced?
A: The lifespan of electrochemical sensors is typically 2-3 years. Actual lifespan depends on the ozone concentration and temperature/humidity in the operating environment, as high-concentration environments accelerate sensor aging.
Q: How should I clean and store the instrument?
A: Use a soft, damp cloth to wipe the exterior. Do not use organic solvents (e.g., alcohol, acetone) to clean the sensor area. Store the instrument in a dry, cool place, avoiding direct sunlight and extreme temperatures.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Acetylene (C2H2). Range: 0-100%LEL Resolution: 0.1LEL. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: PID Gas: Methyl Bromide (CH3Br). Range: 0 to 50/100/200/500/1000/2000/3000ppm Resolution: 0.01ppm;0.1ppm;1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle:Electrochemistry Gas: ozone (O3) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: Chlorine Dioxide/CLO2 Range: 0-10ppm;0-1ppm. Resolution: 0.01PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle:Electrochemistry Gas: Formaldehyde (CH2O) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: N2 Range: 0-100%vol Resolution: 1%vol. Accuracy: ≤±2% F.S.

Category: Products Portable gas detector Portable gas detector Key words: Products Model: HNAG920 Measurement object: CO2/Ex/CH4/CO/H2S/O2/N2/NH3/CL2/H2/He/Ar/CH2O/O3/NO/NO2/SO2/C2H4O/PH3/COCL2/HBR/HCL/HCN/CLO2/SIH4/CH3BR/SO2F2/F2/HF/C2H2/CS2/SF6/C2H...

Detection gas:EX_Flammable-CH4 Range:0-100%LEL (3-100%LEL) / 0-5%VOL Detection Principle:Infrared Resolution:0.1%LEL_0.001%VOL Detection Accuracy:±2%F.S Model:SK/MIC-600

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920

Category: Products Flue gas pretreatment system Simple Pretreatment Online Monitoring System Key words: Products Model: SK/MIC-7500 Measurement object: CO2/Ex/CH4/CO/H2S/O2/N2/NH3/CL2/H2/He/Ar/CH2O/O3/NO/NO2/SO2/C2H4O/PH3/COCL2/HBR/HCL/HCN/CLO2/SIH4/...

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle: Thermal conductivity Gas: helium (He) Range: 0-100%vol Resolution: 1%vol. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle:Electrochemistry Gas: ozone (O3) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: Hydrogen Fluoride (HF) Range: 0-10ppm,0-50ppm,0-100ppm Resolution: 0.01ppm,0.1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Electrochemistry Gas: Ethanol(C2H6O) Range: 0-10ppm,0-50ppm,0-100ppm,Custom Resolution: 0.01ppm,0.1ppm,1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Electrochemistry Gas: Ethylene (C2H4) Range: 0-10ppm,0-50ppm,0-100ppm,0-200ppm. Resolution: 0.01ppm,0.1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Acetylene (C2H2). Range: 0-100%LEL Resolution: 0.1LEL. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: PID Gas: Methyl Bromide (CH3Br). Range: 0 to 50/100/200/500/1000/2000/3000ppm Resolution: 0.01ppm;0.1ppm;1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle:Electrochemistry Gas: ozone (O3) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: Chlorine Dioxide/CLO2 Range: 0-10ppm;0-1ppm. Resolution: 0.01PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle:Electrochemistry Gas: Formaldehyde (CH2O) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: N2 Range: 0-100%vol Resolution: 1%vol. Accuracy: ≤±2% F.S.

Category: Products Portable gas detector Portable gas detector Key words: Products Model: HNAG920 Measurement object: CO2/Ex/CH4/CO/H2S/O2/N2/NH3/CL2/H2/He/Ar/CH2O/O3/NO/NO2/SO2/C2H4O/PH3/COCL2/HBR/HCL/HCN/CLO2/SIH4/CH3BR/SO2F2/F2/HF/C2H2/CS2/SF6/C2H...

Detection gas:EX_Flammable-CH4 Range:0-100%LEL (3-100%LEL) / 0-5%VOL Detection Principle:Infrared Resolution:0.1%LEL_0.001%VOL Detection Accuracy:±2%F.S Model:SK/MIC-600

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920