Category: Products Flue gas pretreatment system On-line monitoring system for high temperature pretreatment
Key words: Products
Model: SK-7500-H2S
Measurement object: Hydrogen sulfide
Category: Products Flue gas pretreatment system On-line monitoring system for high temperature pretreatment
Key words: Products
Model: SK-7500-H2S
Measurement object: Hydrogen sulfide

Function:
The sampled gas first passes through the filter, filters dust and impurities, and then passes through the sampling pump to the electronic condenser (the electronic condenser is mainly used for cooling and dehumidification to avoid secondary condensation of water vapor due to temperature, semiconductor condensation dehumidification, and condensation of humidity water vapor into water for discharge). The flow rate of the extracted gas is controlled by the flow meter (the flow rate of the flow meter is generally controlled at 0.3~0.5L/Min), Finally, in the instrument detection (the working environment of the instrument needs to be used in the working condition of -20 ° ~ 60 °), the water produced by the filter can be automatically drained by the peristaltic pump.
Consumables:
The service life of the sensor is about 1 to 3 years, which is a consumable, and the specific use time is determined according to the detection environment.
2. The color of the chip inside the filter changes and needs to be replaced. The service life is about 3 months, depending on the detection environment.
3. The sampling air pump has a service life of about 1 year and the peristaltic pump has a service life of about 1 year.
Features:
1. Detector detection principle: electrochemical, imported sensors, using intrinsic safety technology, can support multi-gas, multi-range detection, and can provide customized products according to user needs, without tools to achieve sensor exchange, offline calibration and zero self-calibration
2. A variety of signal output, which can be easily connected to PLC/DCS and other industrial control systems, can also be used as a stand-alone control (switch output, RS485 output, 4-20mA output)
3. Local alarm indication, integrated sound and light detector (optional)
4. The instrument has over-range and reverse polarity protection, which can avoid the danger caused by improper operation
5. Cooling and dehumidification of condenser to avoid secondary condensation of water vapor
6. Peristaltic pump with time relay timing drainage
7. Dust filter, filter dust, when there is water, stored with peristaltic pump for discharge
8. Sampling pump, the longest sampling distance can reach 100 meters, it is recommended to be within 20 meters
9. Flow meter, control the gas flow rate, avoid the gas flow too fast to wash the sensor
10. Seven-inch LCD large screen, 1000 historical alarm records, 100000 historical data, adjustable historical data storage interval
Product parameters:
Display mode: 7 inch touch color screen display
Working environment:- 40 ℃ ~ +70 ℃,≤ 95% RH
Sampling temperature:- 40 ℃ ~ 200 ℃ (standard), optional:- 40 ℃ ~ 400 ℃, -40 ℃ ~ 800 ℃, -40 ℃ ~ 1300 ℃
Sampling humidity: 0 ~ 99% RH
Detection range: 0-10/100/200/500/1000/2000/4000/10000/20000ppm optional
Output signal: 4 ~ 20mA, RS485, multiple passive contacts, optional: DTU wireless transmission, network transmission and other ways
Constant temperature: 25 ℃ temperature after gas treatment, can be set
Constant humidity: 70% RH, humidity after gas treatment, automatic
Automatic drainage: according to the size of the water vapor automatic drainage
System heating: optional function to prevent freezing of pipelines in extremely cold areas
Working mode: fixed continuous work, pump suction measurement
Sampling flow rate: 4 l/min (standard)
Working voltage: 220VAC,50HZ,200W
Overall dimensions: 700CMX500CMX270CM (see the appearance size chart)
Protection level: IP65 waterproof type (optional outdoor use)
Explosion-proof outer box: optional


Function:
The sampled gas first passes through the filter, filters dust and impurities, and then passes through the sampling pump to the electronic condenser (the electronic condenser is mainly used for cooling and dehumidification to avoid secondary condensation of water vapor due to temperature, semiconductor condensation dehumidification, and condensation of humidity water vapor into water for discharge). The flow rate of the extracted gas is controlled by the flow meter (the flow rate of the flow meter is generally controlled at 0.3~0.5L/Min), Finally, in the instrument detection (the working environment of the instrument needs to be used in the working condition of -20 ° ~ 60 °), the water produced by the filter can be automatically drained by the peristaltic pump.
Consumables:
The service life of the sensor is about 1 to 3 years, which is a consumable, and the specific use time is determined according to the detection environment.
2. The color of the chip inside the filter changes and needs to be replaced. The service life is about 3 months, depending on the detection environment.
3. The sampling air pump has a service life of about 1 year and the peristaltic pump has a service life of about 1 year.
Features:
1. Detector detection principle: electrochemical, imported sensors, using intrinsic safety technology, can support multi-gas, multi-range detection, and can provide customized products according to user needs, without tools to achieve sensor exchange, offline calibration and zero self-calibration
2. A variety of signal output, which can be easily connected to PLC/DCS and other industrial control systems, can also be used as a stand-alone control (switch output, RS485 output, 4-20mA output)
3. Local alarm indication, integrated sound and light detector (optional)
4. The instrument has over-range and reverse polarity protection, which can avoid the danger caused by improper operation
5. Cooling and dehumidification of condenser to avoid secondary condensation of water vapor
6. Peristaltic pump with time relay timing drainage
7. Dust filter, filter dust, when there is water, stored with peristaltic pump for discharge
8. Sampling pump, the longest sampling distance can reach 100 meters, it is recommended to be within 20 meters
9. Flow meter, control the gas flow rate, avoid the gas flow too fast to wash the sensor
10. Seven-inch LCD large screen, 1000 historical alarm records, 100000 historical data, adjustable historical data storage interval
Product parameters:
Display mode: 7 inch touch color screen display
Working environment:- 40 ℃ ~ +70 ℃,≤ 95% RH
Sampling temperature:- 40 ℃ ~ 200 ℃ (standard), optional:- 40 ℃ ~ 400 ℃, -40 ℃ ~ 800 ℃, -40 ℃ ~ 1300 ℃
Sampling humidity: 0 ~ 99% RH
Detection range: 0-10/100/200/500/1000/2000/4000/10000/20000ppm optional
Output signal: 4 ~ 20mA, RS485, multiple passive contacts, optional: DTU wireless transmission, network transmission and other ways
Constant temperature: 25 ℃ temperature after gas treatment, can be set
Constant humidity: 70% RH, humidity after gas treatment, automatic
Automatic drainage: according to the size of the water vapor automatic drainage
System heating: optional function to prevent freezing of pipelines in extremely cold areas
Working mode: fixed continuous work, pump suction measurement
Sampling flow rate: 4 l/min (standard)
Working voltage: 220VAC,50HZ,200W
Overall dimensions: 700CMX500CMX270CM (see the appearance size chart)
Protection level: IP65 waterproof type (optional outdoor use)
Explosion-proof outer box: optional


Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Acetylene (C2H2). Range: 0-100%LEL Resolution: 0.1LEL. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: PID Gas: Methyl Bromide (CH3Br). Range: 0 to 50/100/200/500/1000/2000/3000ppm Resolution: 0.01ppm;0.1ppm;1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle:Electrochemistry Gas: ozone (O3) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: Chlorine Dioxide/CLO2 Range: 0-10ppm;0-1ppm. Resolution: 0.01PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle:Electrochemistry Gas: Formaldehyde (CH2O) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: N2 Range: 0-100%vol Resolution: 1%vol. Accuracy: ≤±2% F.S.

Category: Products Portable gas detector Portable gas detector Key words: Products Model: HNAG920 Measurement object: CO2/Ex/CH4/CO/H2S/O2/N2/NH3/CL2/H2/He/Ar/CH2O/O3/NO/NO2/SO2/C2H4O/PH3/COCL2/HBR/HCL/HCN/CLO2/SIH4/CH3BR/SO2F2/F2/HF/C2H2/CS2/SF6/C2H...

Detection gas:EX_Flammable-CH4 Range:0-100%LEL (3-100%LEL) / 0-5%VOL Detection Principle:Infrared Resolution:0.1%LEL_0.001%VOL Detection Accuracy:±2%F.S Model:SK/MIC-600

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920

Category: Products Flue gas pretreatment system Simple Pretreatment Online Monitoring System Key words: Products Model: SK/MIC-7500 Measurement object: CO2/Ex/CH4/CO/H2S/O2/N2/NH3/CL2/H2/He/Ar/CH2O/O3/NO/NO2/SO2/C2H4O/PH3/COCL2/HBR/HCL/HCN/CLO2/SIH4/...

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle: Thermal conductivity Gas: helium (He) Range: 0-100%vol Resolution: 1%vol. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle:Electrochemistry Gas: ozone (O3) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: Hydrogen Fluoride (HF) Range: 0-10ppm,0-50ppm,0-100ppm Resolution: 0.01ppm,0.1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Electrochemistry Gas: Ethanol(C2H6O) Range: 0-10ppm,0-50ppm,0-100ppm,Custom Resolution: 0.01ppm,0.1ppm,1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Electrochemistry Gas: Ethylene (C2H4) Range: 0-10ppm,0-50ppm,0-100ppm,0-200ppm. Resolution: 0.01ppm,0.1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Acetylene (C2H2). Range: 0-100%LEL Resolution: 0.1LEL. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle: PID Gas: Methyl Bromide (CH3Br). Range: 0 to 50/100/200/500/1000/2000/3000ppm Resolution: 0.01ppm;0.1ppm;1ppm. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction and Diffusion Detection Principle:Electrochemistry Gas: ozone (O3) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction and Diffusion Detection Principle: Infrared NDIR Gas: Sulfur hexafluoride (SF6). Range: 0-1000ppm,0-3000PPM,0-5000PPM. Resolution: 1PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: Chlorine Dioxide/CLO2 Range: 0-10ppm;0-1ppm. Resolution: 0.01PPM. Accuracy: ≤±2% F.S.

Model: HNAG900 Sampling Method: Pumping Suction Detection Principle:Electrochemistry Gas: Formaldehyde (CH2O) Range: 0-50PPM,0-100PPM. Resolution: 0.01PPM,0.1PPM. Accuracy: ≤±2% F.S.

Model: HNAG920 Sampling Method: Pumping Suction Detection Principle: Electrochemistry Gas: N2 Range: 0-100%vol Resolution: 1%vol. Accuracy: ≤±2% F.S.

Category: Products Portable gas detector Portable gas detector Key words: Products Model: HNAG920 Measurement object: CO2/Ex/CH4/CO/H2S/O2/N2/NH3/CL2/H2/He/Ar/CH2O/O3/NO/NO2/SO2/C2H4O/PH3/COCL2/HBR/HCL/HCN/CLO2/SIH4/CH3BR/SO2F2/F2/HF/C2H2/CS2/SF6/C2H...

Detection gas:EX_Flammable-CH4 Range:0-100%LEL (3-100%LEL) / 0-5%VOL Detection Principle:Infrared Resolution:0.1%LEL_0.001%VOL Detection Accuracy:±2%F.S Model:SK/MIC-600

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:CO,Carbon monoxide Range:0-1000ppm/0-5000ppm/custom Detection Principle:Electrochemistry Resolution:1ppm Detection Accuracy:±2%F.S Model:HNAG1000

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920

Detection gas:O2/Oxygen Range:0-30%VOL (0-25%VOL) / 0-100%VOL Detection Principle:Electrochemistry Resolution:0.1%VOL/0.01%VOL Detection Accuracy:±2%F.S Model:HNAG920